3.34. Calibrate Stage Option Greyed Out
Why Is the "Calibrate Stage" Option Greyed Out Even Though My X-Y Stage Is Recognized?
The preferences for the X-Y stage are stored in two locations within Volocity:
-
Device Preferences:
- Accessible via Edit > Preferences > Devices (or Volocity > Preferences > Devices on macOS).
-
X-Y Stage Preferences:
- Accessible via Edit > Preferences > X-Y Stage (or Volocity > Preferences > X-Y Stage on macOS).
Troubleshooting Steps:
- Verify that the stage appears in the Device Preferences and that the X-Y Stage option is selected.
- If the "Calibrate Stage" function in the Stage menu is still greyed out, ensure that an appropriate X-Y stage device is selected in the X-Y Stage Preferences.
This ensures that Volocity recognizes the correct stage settings and enables the calibration functionality.